Franssen Pascal M L - - 2010
We report on the validation of the new generation Mobil-O-Graph 24/48 h ambulatory blood pressure monitor according to the criteria of the European Society of Hypertension. In 15 individuals participating in phase I for systolic pressure, all 45 measures differed less than 15 mmHg, 43 and 33 out of 45 ...
Plotkin Anton - - 2010
Paradoxically, improvements in emergency medicine have increased survival albeit with severe disability ranging from quadriplegia to "locked-in syndrome." Locked-in syndrome is characterized by intact cognition yet complete paralysis, and hence these individuals are "locked-in" their own body, at best able to communicate using eye blinks alone. Sniffing is a precise ...
Li Liang L Department of Chemistry and Institute for Biophysical Dynamics, The University of Chicago, 929 East 57th Street, Chicago, Illinois 60637, - - 2010
In this paper, we describe a method to load a microfluidic device, the SlipChip, via dead-end filling. In dead-end filling, the lubricating fluid that fills the SlipChip after assembly is dissipated through the gap between the two plates of the SlipChip instead of flowing through an outlet at the end ...
Lee Myeong Soo - - 2010
The aims of this study were to develop a sham cupping device and to validate its use as a placebo control for healthy volunteers. A sham cupping device was developed by establishing a small hole to reduce the negative pressure after suction such that inner pressure could not be maintained ...
Cecconi M - - 2010
Pulse pressure (PP) analysis from a radial arterial line is available with the LiDCO plus monitor (LiDCO, Cambridge, UK) and FloTrac/Vigileo (Edwards Lifesciences, Irvine, CA, USA). The aim of this study was to investigate the agreement of continuous PP analysis against intermittent thermodilution (ITD) using the pulmonary artery catheter (PAC). ...
Yoshikawa Mineka - - 2011
A new tongue pressure device consisting of a simple and safe disposable probe and manometer has been developed. This report describes a study that examined the validity of the new device, comparing it to a widely used tongue pressure manometer, the Iowa Oral Performance Instrument (IOPI), and to the stable ...
Petric Aleksandra Domanjko - - 2010
Indirect blood pressure measurements using high definition oscillometric (HDO) and Doppler devices were compared in 50 anaesthetised client-owned cats presented for various surgical procedures. Sites of cuff placement for Doppler were identified as forelimb and hindlimb and for HDO as forelimb and tail. Oscillometric and Doppler readings were obtained in ...
Mantke René - - 2011
Experimental data about the efficacy and safety of sealing devices are rare. Therefore, this study investigated these parameters for three commercially available energy-based vascular sealing and cutting systems. In male hybrid pigs, 487 carotid artery segments were sealed and cut using the harmonic scalpel or several bipolar sealing devices. The ...
Pérez-Martin A - - 2010
This study was designed to assess the accuracy and reliability of a new, portable, fully automated photoplethysmography (PPG) device for toe blood pressure (TBP) measurement. Within-subject comparison with conventional laser Doppler (LD) measurement. Four TBP measurements were performed on both lower limbs, alternatively with LD and PPG in 200 patients ...
Gräser Oliver - - 2010
An algorithm for an adaptive boundary condition for the simulation of fully developed, pressure driven flows using the lattice-Boltzmann method is introduced. The method simulates the effect of the solid walls confining a structure with dual periodicity, such as used in microfluidic devices. We combine a recently introduced boundary condition ...
Suh Lara K - - 2010
INTRODUCTION: Pressurized saline irrigation is commonly used during ureteroscopy, which can cause an increase in intrarenal pressure leading to postoperative pain, sepsis, and renal injury due to pyelovenous and pyelolymphatic backflow. To prevent retrograde stone migration during ureteroscopic lithotripsy, antiretropulsion devices can be deployed, which may or may not protect ...
Agrawal Megha - - 2010
Persistence of the ductus arteriosus (DA), connecting the aorta to the pulmonary artery, after birth leads to a congenital heart disease known as Patent Ductus Arteriosus (PDA). The objective of this study was to propose a new intraductal design for PDA occlusion and determine the efficacy of the new device ...
Toh Gwendoline M - - 2010
In this article, we report the design of a microchip based hydraulic pump that employs a sodium silicate derived sol-gel structure for generating pressure-driven flow within a microfluidic network. The reported sol-gel structure was fabricated in a chosen location of our device by selectively retaining sodium silicate solution within a ...
Coleman Andrew - - 2010
BACKGROUND: The BP11 is currently one of the lowest-cost blood pressure (BP) monitors available in the UK and has sold approximately 1.5 million units since its launch in 2006. As access to home-use BP monitoring increases, it is important that such devices are validated so that consumers and clinicians can ...
Pan Feng - - 2010
OBJECTIVE: Blood pressure measuring devices should be subjected to an independent evaluation of accuracy before clinical use. The results of the validation of the Andon KD-575 automated blood pressure monitor according to the European Society of Hypertension International Protocol in adults are presented in this paper. METHODS: Sequential same-arm measurements ...
Muecke Sandy - - 2010
OBJECTIVE: Non-invasive blood pressure recordings may be inaccurate in the critically ill patient and measurement difficulties are intensified in the prehospital setting. This may adversely impact upon outcomes for many critically ill patients, particularly those with traumatic brain injury and/or lengthy prehospital times. This study aimed to validate a non-invasive, ...
Broelmann Hannah - - 2010
OBJECTIVE: The validation of two new oscillometric blood pressure (BP) measurement devices according to the International Protocol (European Society of Hypertension). METHODS: Two trained observers measured BP in 33 patients with nine sequential same-arm measurements each, whereby five measurements with the mercury standard device alternated with four measurements with each ...
Pham Thuy - - 2010
The coronary sinus (CS) vessel serves as a conduit for the deployment of percutaneous transvenous mitral annuloplasty (PTMA) devices for the treatment of functional mitral regurgitation. Characterization of the mechanical response of the CS is an important step towards an understanding of tissue-device interaction in PTMA intervention. The purpose of ...
McCormick Daniel - - 2010
1. Measurement of blood pressure via telemetry in a variety of animal species has become an indispensable part of cardiovascular physiology, drug development and safety pharmacology. 2. These telemetry systems use fluid-filled catheters, which differ from commonly encountered indwelling catheters by virtue of their short length, high durometer, variable inner ...
Genzwuerker H V - - 2010
AIM: This study investigates if a n impedance threshold valve (ITV) might improve survival after cardiac arrest by increasing vital organ blood flow. The combination of ITV and supraglottic airway devices (SADs) has not been previously studied. This simulation study in a manikin aimed at analysing differences in ventilation with ...
Wernick M - - 2010
OBJECTIVES: To evaluate high-definition and conventional oscillometry in comparison with direct blood pressure measurements in anaesthetised dogs. METHODS: Eight simultaneous readings for systolic, diastolic and mean pressure were obtained directly and with each of two devices in nine anaesthetised dogs. Measurement procedure and validation were based on the 2007 ACVIM ...
Dixon M J - - 2010
Air pressure is commonly used to drive a mechanical stimulus for nociceptive threshold testing. This may be bulky, noisy, non-linear and suffer from friction, hence development of a better system is described. A novel, light (14 g) rolling diaphragm actuator was constructed, which supplied 20 N force via a constant ...
Gaddum Nicholas Richard - - 2010
Clinical studies have reported the balancing of pump outputs to be a serious control issue for rotary biventricular support (BiVS) systems. Poor reliability of long-term, blood immersed pressure sensors encouraged the development of a new control strategy to improve their viability. A rotary BiVS device was designed and constructed with ...
Aliño M - - 2010
The influence of sodium chloride on the water holding capacity of meat is well known and described in literature. An increase in water holding capacity is associated with an increase in the swelling pressure in the protein matrix. A novel device was used to measure the swelling pressure during brine ...
Myllylä Teemu S - - 2011
This report focuses on designing and implementing a non-invasive blood pressure (NIBP) measuring device capable of being used during magnetic resonance imaging (MRI). Based on measuring pulse wave velocity in arterial blood, the device uses the obtained result to estimate diastolic blood pressure. Pulse transit times are measured by two ...
Giacomozzi Claudia - - 2010
Accurate plantar pressure measurements are mandatory in both clinical and research contexts. Differences in accuracy, precision and reliability of the available devices have prevented so far the onset of standardization processes or the definition of reliable reference datasets. In order to comparatively assess the appropriateness of the most used pressure ...
Wan Y Y Department of Health Statistics, Fourth Military Medical University, Xi'an, - - 2010
The aim of this study is to systematically examine the proportion of accurate readings attained by automatic digital blood pressure (BP) devices in published validation studies. We included studies of automatic digital BP devices using recognized protocols. We summarized the data as mean and s.d. of differences between measured and ...
Kulkarni Atul - - 2010
Elastomer has become a material of much interest for use as a deformation element in pressure and force monitoring devices. In the present work, we fabricated and characterized a pressure sensor that uses the polydimethylsiloxane (PDMS) elastomer and the plastic optical fiber (POF). The POF is used to guide light ...
Abe J - - 2010
A compact cubic-anvil high-pressure device was developed for in situ neutron powder diffraction studies. In this device, a cubic shaped pressure medium is compressed by six anvils, and neutron beams pass through gaps between the anvils. The first high-pressure experiment using this device was conducted at J-PARC and clearly showed ...
Yu Yi - - 2010
Brain-machine interface (BMI) has been developed due to its possibility to cure severe body paralysis. This technology has been used to realize the direct control of prosthetic devices, such as robot arms, computer cursors, and paralyzed muscles. A variety of neural decoding algorithms have been designed to explore relationships between ...
Amoore J N - - 2010
Patient management requires accurate blood pressure measurement, but there is concern about the accuracy of the equipment. Makes and types of sphygmomanometers and sizes of their associated cuffs used in hospital wards and outpatient departments were recorded and calibration checked. Half the sphygmomanometers were oscillometric, 32% aneroid and 18% mercury. ...
Kruger Diane M - - 2010
The LongPen is a remote-controlled pen and videoconferencing device conceived by Canadian author Margaret Atwood in 2004 and initially intended to bring "live" author signings to far away locations. The LongPen allows for individually inscribed long distance signatures and writing while maintaining an original record, written with pen and ink. ...
Asmar Roland - - 2010
OBJECTIVE: Three professional devices for office blood pressure (BP) measurement, using three different algorithms to determine BP, were evaluated according to the International Protocol of the European Society of Hypertension. The Omron BP10 uses the oscillometric method, the Omron HBP T105 (module HBP-M3600) uses the smart inflation mode and high-speed ...
Hemmes Baukje - - 2010
BACKGROUND:: Immobilization of the spine in patients with trauma at risk of spinal damage is performed using a rigid long spineboard or vacuum mattress both during prehospital and in-hospital care. However, disadvantages of these immobilization devices in terms of discomfort and tissue-interface pressures have guided the development of a new ...
Powner David J - - 2010
Many biological signals must be measured and interpreted accurately to titrate therapy properly during donor care. Although the technological aspects of intravascular pressure monitoring are usually delegated to bedside nursing colleagues, organ procurement coordinators should be familiar with those devices and methods. The equipment, supplies, and procedures used for arterial ...
Asmar Roland - - 2010
OBJECTIVE: Three oscillometric devices for the self blood pressure measurement were evaluated according to the International Protocol of the European Society of Hypertension in three separate studies. The Omron M3 Intellisense and the Omron M2 Compact measures blood pressure (BP) at the brachial level; the Omron R3-I Plus measures BP ...
Bonso Elisa - - 2010
To evaluate the accuracy of the UEBE Visomat Comfort form device for home blood pressure measurement tested following the recommendations of the International Protocol of the European Society of Hypertension. Device evaluation was performed in 33 participants with a mean +/- standard deviation age of 56.5+/-18.3 years (range 30-91 years). ...
Stergiou George S - - 2010
OBJECTIVE: To perform a systematic review of validation studies of blood pressure measuring devices done using the European Society of Hypertension International Protocol (ESH-IP) since its publication in 2002. METHODS: Major aspects of ESH-IP studies are described. A review of the ESH-IP performance, problems and violations in its application, and ...
Tasker F - - 2010
Both automated and auscultatory blood pressure (BP) devices have their strengths and accuracy limitations. Hybrid devices, such as the Nissei DM-3000, are mercury free and provide both automated and auscultatory measurement modes. The aim of this study was to validate all measurement modes of the Nissei DM-3000 device according to ...
Guzelsu N - - 2010
A novel streaming potential measurement device has been validated by determining the average electrokinetic (zeta) potential of densely packed particulate such as human erythrocytes and ground bovine cortical bone. The new streaming potential device used in this study is easy to construct in the laboratory, designed to allow dense packing ...
Butler Kenneth R - - 2010
The objective of this study was to evaluate terminal digit preference in blood pressure (BP) measurements taken from a sample of clinics at a large academic health sciences center. We hypothesized that terminal digit preference would occur more frequently in BP measurements taken with manual mercury sphygmomanometry compared to those ...
Khaliq F - - 2010
Medical training is highly stressful particularly for those who are beginning their medical education. The present study was undertaken to find out the level of stress, hypertension and its etiological factors among undergraduate medical students. One hundred and seventeen undergraduate medical students between 17-21 years of age got examined for ...
Riscica Fabio - - 2010
In recent years, the usefulness of High Frequency Percussive Ventilation (HFPV) has been clinically valued as a suitable alternative to conventional mechanical ventilation especially in particularly conditions (for example in pulmonary resection [1]). However, sophisticated and expensive equipments are required for measuring the respiratory parameters during HFPV, because of the ...
Jiang Guangqiang - - 2010
Pressure in various organs and body parts, such as blood vessels, heart, brain, eyes, bladder and GI tracts, is an important indication of health. Long term, continuous pressure monitoring is critically needed for a number of applications. When combined with existing neuro-prosthetics devices, they may provide better solutions to many ...
de Greeff A - - 2010
Accurate blood pressure (BP) measurement is dependent on a trained observer using validated and properly maintained equipment. BP devices should be checked regularly to ensure that their calibration remains within the European Standard specification of +/-3 mm Hg. This study assessed the air leakage rates and calibration accuracy of BP ...
Trevino Rodolfo J - - 2010
Abstract The Scisense (London, ON, Canada) micro-manometer pressure sensor is currently being used by investigators to evaluate cardiovascular physiology in mice, but has not been validated to date. The purpose of the current study is to compare the 1.2 F Scisense pressure sensor to the current gold standard produced by ...
Iff Isabelle - - 2010
OBJECTIVE: To determine the usefulness of an acoustic device to confirm correct placement of extradural needles in horses. STUDY DESIGN: Prospective experimental study. ANIMALS: Twelve adult healthy horses weighing between 434 and 640 kg. MATERIALS AND METHODS: Horses were sedated, placed in stocks and subjected to caudal extradural anaesthesia using ...
Nakamura Kaori - - 2010
To validate the FM-800 device (Fukuda Denshi Co., Ltd., Tokyo, Japan) for ambulatory blood pressure monitoring (ABPM), this validation study was performed based on the European Society of Hypertension (ESH) protocol and the Association for the Advancement of Medical Instrumentation (AAMI) criteria. Thirty-three participants were included (15 in phase 1 ...
Knops Simon P - - 2010
Data on the efficacy and safety of non-invasive Pelvic Circumferential Compression Devices (PCCDs) is limited. Tissue damage may occur if a continuous pressure on the skin exceeding 9.3 kPa is sustained for more than two or three hours. The aim of this study was to gain insight into the pressure ...
Coleman Tanner J TJ Department of Bioengineering, University of Utah, Salt Lake City, UT84112, USA. - - 2010
Limitations of the standard urogynecological pressure transducers have not adequately provided reliable measurements of intra-abdominal pressure (IAP) during physical activities. A previous novel intravaginal pressure transducer (IVT) was developed in order to overcome the shortcomings in existing technology. The design of the IVT was validated though comparisons with existing pressure ...