Effectiveness of resonance frequency in predicting orthopedic implant strength and stability in an in vitro osseointegration model.
Article Type: Report
Subject: Orthopedic implants (Design and construction)
Orthopedic implants (Health aspects)
Rehabilitation (Research)
Authors: Isaacson, Brad M.
Vance, Richard E.
Chou, Teri G. Rosenbaum
Bloebaum, Roy D.
Bachus, Kent N.
Webster, Joseph B.
Pub Date: 12/31/2009
Publication: Name: Journal of Rehabilitation Research & Development Publisher: Department of Veterans Affairs Audience: Academic Format: Magazine/Journal Subject: Health Copyright: COPYRIGHT 2009 Department of Veterans Affairs ISSN: 0748-7711
Issue: Date: Dec 31, 2009 Source Volume: 46 Source Issue: 9
Topic: Event Code: 310 Science & research
Geographic: Geographic Scope: United States Geographic Code: 1USA United States
Accession Number: 227281563
Full Text: INTRODUCTION

The term "osseointegration" has been used to describe a structural and functional connection between living bone and the surface of a load bearing implant [1-2]. Successful osseointegration techniques have been previously demonstrated for individuals with limb loss [3-5], as well as in the areas of dentistry [1-2,6-8] and craniofacial reconstruction [6]. As a result of an increased desire for functionality for patients with transfemoral, transhumeral, and transtibial amputations [3,9-11], osseointegration technology has been developed for direct skeletal attachment of an exoprosthesis to the residual limb. With the osseointegration procedure for persons with limb loss, a biocompatible metal fixation is surgically inserted directly into the bone of the residual limb and serves as an attachment system for connecting and suspending a prosthesis to the residual limb [5,12-13].

One challenge with the use of natural biological skeletal fixation is allowing the bone to heal and osseointegrate with the implant surface, thereby attaining a strong skeletal interlock, a prerequisite for long-term implant function and stability [14-15]. To prevent mechanical loosening at the bone-implant construct, current osseointegration rehabilitation protocols require extensive periods of restricted load bearing to allow sufficient bone attachment and prevent overloading at the bone-implant interface [12,16-24]. Limiting the force on the periprosthetic bone following insertion of an osseointegrated implant is based on the principle that stress must be exerted gradually to promote firm skeletal attachment, since under- or overloading may compromise the integrity of the host bone. However, previous literature indicates that immediate implant loading may not compromise the integrity of the bone-implant interface or prevent osseointegration if micromotion is controlled in properly designed implants [15,17-19,23-27].

The inability to quantitatively diagnose implant fixation has driven scientists and physicians alike to develop clinical tools to assess primary and secondary implant stability. Primary implant stability occurs immediately after surgical placement [28], and successful osseointegrated implants result from proper implant fit and fill [29] and surgical techniques [30-32]. However, secondary implant stability is the result of bone healing and remodeling that occurs over time [33]. While initial implant fixation is required to prevent micromotion and fibrous encapsulation [16,19,34-40], the long-term success of osseointegrated implants requires firm skeletal connections that may not occur until 9 months postoperative in human cancellous bone [16] and would be more representative of both primary and secondary implant stability.

Since no quantitative measures exist to determine optimal skeletal attachment in vivo, current European and Australian rehabilitation programs for transfemoral amputees with osseointegrated implants require slow, progressive weight bearing [41-42]. Methods for assessing osseointegration strength and stability have included mechanical testing, light microscopy, and scanning electron microscopy [43-46]; however, these techniques for evaluating skeletal fixation cannot be used in dental applications and for patients with limb loss because of the invasiveness [46-47]. Therefore, alternative means of assessing implant stability have been developed and include radiographs, percussion tests, and resonance frequency. Radiographs, however, are not sensitive for determining the extent of osseointegration [48] since a reduction in bone mineralization of 40 percent [49] is required before bone mineral loss can be accurately determined. An examination of radiolucencies with plain radiographs has shown to result in approximately 2 percent false-positive diagnoses of inadequate osseointegration and implant loosening [50]. In addition, traditional radiographs require standardization with threaded implants [21], since reproducibility is difficult and essential in preventing unnecessary surgical revisions [50]. Advanced imaging techniques such as computed tomography scans are also not practical because of expenses, image artifact generated from metal implants, and high exposure to radiation [51].

Percussion tests have also been used to assess osseointegration implant stability, but this technique is considered generally inadequate in the clinical setting. When implant stability is evaluated with percussion tests using dental implants, the procedure often results in "more information about the tapping instrument, and will at best only reveal poor qualitative information" [52]. Therefore, percussion tests are limited since the process only provides quick distinction between mobile and osseointegrated implants but does not reveal the degree of implant stability and thereby restricts the ability of rehabilitation specialist to monitor and advance progressive weight-bearing regimens [53].

A potentially useful noninvasive alternative is resonance frequency analysis (RFA), which has been shown to correlate with the degree of implant stability in dental implants [21,44,47-48,52-55]. Fluctuations in implant stability quotient (ISQ) values even mimic expected bone remodeling rates [20,54,56]. Several studies have also reported that ISQ values decrease within the first 3 months postoperative [20,57-58], as a result of osteoclast resorption, which is key to increased osteoblast activity and new bone formation [33,59]. The reliability of resonance frequency has been confirmed with ISQ values but is dependent on jaw position [44,53] and attributed to the differences in bone quality and type [17, 37-38,43,60].

The resonance frequency system most commonly used with dental implants and reported in the literature is the Osstell Mentor[R] (Osstell AB; Gothenburg, Sweden). This device uses a compact battery system and magnetic pulse to monitor ISQ values [61]. The handheld probe is attached to a docking station, and a magnetic pulse is transmitted to a Smartpeg (Osstell AB) temporarily attached to the proximal face of the implant [41,44]. With use of a bode plot function [61], the response signal is translated into a numerical ISQ value, which may be used to determine the degree of implant stability. Extensive review articles evaluating the performance of the Osstell device are also available but are not within the scope of this narrative [49,52]. While RFA has been shown to effectively determine osseointegration implant stability in edentulous mandible and maxilla models, ours is the first study to evaluate the utility of this device in an in vitro model simulating irregular osseointegration in a long bone medullary canal.

The objective of this study was to determine a biomechanical relationship between RFA ISQ values and the force required to displace a titanium alloy implant. A composite fixture was utilized to improve reproducibility, which arises from the use of a more uniform material type. Establishing a relationship between torque, attachment location, and resonance frequency with mechanical pushout forces would provide valuable information as a basis for further in vivo testing, because currently no biomechanical testing data directly relate resonance frequency to load bearing. Clearly, new developments in orthopedics and rehabilitation will continue to demand advanced tools for patients seeking osseointegrated implants following limb loss. Therefore, the ultimate goal of this research is to utilize resonance frequency for optimizing the speed and safety of rehabilitation for amputees with osseointegrated implants, since skeletal fixation will vary between persons [49], and individual rehabilitation programs will be required.

METHODS

Testing Fixture and Implant Design

An implant fixation construct was designed to simulate percutaneous, osseointegrated implant attachment in a medullary canal and was built in a three-dimensional printer (Dimension Elite, Stratasys Inc; Eden Prairie, Minnesota) utilizing layered acrylonitrile butadiene styrene (ABS) plastic. The completed fixture was 45 mm in height by 30 mm in width with a 4 mm hole centrally located for implant attachment. An ABS model was selected as the testing material because it excludes the inherent variability of human cadaveric bone [62-64]. In addition to variability between bone specimens, dehydrated bone will exhibit a higher modulus of elasticity caused by the diffusion of water into vacant spaces and stiffening of collagen fibers [65]. This decrease in water content, which may occur during biomechanical testing, will subsequently lead to the host bone becoming brittle and exhibiting reduced plasticity [65] and may lead to inaccurate data collection. Since the protocol was initially targeting the precision and accuracy of the resonance frequency technique for potential use in patients with osseointegrated implants, a standard material type was selected.

Titanium alloy (Ti-6Al-4V) implant analogs, 20 mm in length, were manufactured to investigate the utility of the resonance frequency device. Implants were fabricated as one unit and included a hexagonal head (5 mm in height by 8 mm in width) and a smooth post distally (15 mm in length and 3.5 mm in diameter). The hexagonal design included an 8 mm base to fit securely on the ABS device and prevent translation and rotation during data collection (Figure 1). Titanium alloy was also selected as the implant material based on its frequent successful use in total joint replacements [66-67], biocompatibility [6869], and nonferromagnetic properties [70-71] that would not alter RFA recordings.

Three 3 mm holes were evenly spaced and drilled 5 mm apart in the ABS fixture to represent partial osseointegration along the titanium implant insert (Figure 1). Setscrews were fixed at multiple insertion torques and levels to prevent the smooth titanium implant from migrating and to simulate a series of nonuniform bone-implant interfaces and depths, since the quality of osseointegration cannot be assumed to be uniform along the bone-implant construct. In the ABS fixture, Hole A was located 20 mm from the most proximal aspect of the Smartpeg and each additional hole positioned 5 mm distally. Implants were secured with torques ranging from 5 to 12.5 N x cm, since 5 to 50 N x cm is regarded as necessary for primary implant stability [47]. Therefore, to ensure accurate data collection in this model and avoid exceeding the elastic limit of the setscrews, we chose 12.5 N x cm as the highest evaluated torque. A range of torques was also selected in the in vitro model, since the initial fixation required for successful osseointegration is unknown [47] and depends largely on bone quality and density [72].

[FIGURE 1 OMITTED]

Resonance Frequency Analysis

The Osstell measurement system (Osstell Mentor[R]) designed for oral cavity and craniofacial implants was used to obtain the resonance frequency values. Smartpegs were attached to the titanium alloy implants and received magnetic pulses to determine implant fixation strengths. The threaded Smartpegs (Type 1) were consistently torqued to 10 N x cm with a hand-held, digital torque meter (Advanced Force/Torque Indicator, Dillon Quantrol; Fairmont, Minnesota), and ISQ values were recorded before mechanical testing, based on recommendations from previous RFA literature [47]. An insertion torque of 10 N x cm was also selected since it was regarded as least likely to damage the Smartpeg's screw threads [21] and would ensure accurate data collection.

Skeletal fixation was simulated with six variable torques (5.0, 7.0, 8.6, 10.0, 10.9, and 12.5 N x cm) at three separate fixation locations (Hole A, Hole B, and Hole C) (Figure 1). Screw torques were randomly selected within the predetermined range and verified with the digital torque meter before data collection. After recording three ISQ's values on each orthogonal face to ensure implants were properly loaded on all axes, to demonstrate repeatability, and per manufacturer instructions [44], we removed the Smartpegs and positioned the implant below the crosshead of the mechanical testing apparatus (Model 8800, Instron Corp.; Norwood, Massachusetts). Smartpegs were discarded and replaced after three measurements to avoid screw thread wearing.

Mechanical Testing

Implants were secured in the ABS fixture, placed in the servohydraulic mechanical testing device, and preloaded to 30 N in a load-controlled testing mode. Load was increased steadily at a rate of 5 N/s (a recommended loading rate established by the mechanical engineering coauthors) until implant displacement occurred. All data were sampled continuously at 100 Hz throughout the mechanical-testing procedure (Figure 2). To prevent observer bias, we determined the exact point of implant displacement with a custom script (MATLAB, The MathWorks Inc; Natick, Massachusetts). The code was generated to determine a 3 percent change in slope for the continuous data, and the data point that corresponded with the initial change in slope was selected as the ultimate failure load. A 3 percent change in slope between position and time accounted for the specific tolerance of our mechanical testing system, as well as a safety factor of 1 percent. Therefore, a 3 percent relative difference in slope would likely account for measurement error and system noise and prevent subjective interpretations of implant displacement.

To make certain that the implant construct was not damaged from repeated mechanical-testing usage, the ABS fixture was carefully bivalved (Craftsman 10 in. Direct Drive Band Saw, Sears Holding Corporation; Hoffman Estates, Illinois) and examined with a laboratory macroscope (Nikon SMZ800, Nikon Inc; Melville, New York) equipped with imaging capturing software (Magnafire SP, Optronics; Goleta, California) at the conclusion of the pushout tests. Investigating the inner wall of the fixation device was important for ensuring that damage did not occur and alter the surface area for implant attachment or affect the predetermined loading conditions. Extensive examination of the construct did not reveal visible imperfections from mechanical testing and helped demonstrate repeatability and accuracy during data collection.

[FIGURE 2 OMITTED]

Statistical Analysis

A multiple linear regression was used to correlate the outcome of mechanical pushout force with the predictor variables (screw torque, fixation distance, resonance frequency ISQ, interaction between resonance frequency and screw torque, and distance from Smartpeg). In one model, the distance from the Smartpeg was included as a continuous variable; in the second model, it was investigated as two dummy variables to account for the three points of fixation. In each case, the interaction between screw torque and resonance frequency was included. To avoid overfitting, at least 10 observations for each predictor variable were required in the model. In this study with a sample size of 45, four to five variables were included and would not lead to misinterpretation of data. R2 values are reported with each model along with the adjusted R2 values, which represent the correlation without problems due to overfitting. All statistical comparisons were conducted with commercially available software (SPSS, Inc; Chicago, Illinois).

RESULTS

Descriptive statistics (mean and standard deviation [SD]) based on mechanical displacement forces for fixation location and screw torque are provided in Table 1. When determining statistical significance, we excluded torques exceeding 10.9 N x cm, since inconsistent data indicated that the elastic limit of the screws was exceeded and inclusion would have provided speculative correlations. Therefore, a total of 45 data points was collected for three fixation locations and five torques on the single fixture. Further interpretation of Table 1 demonstrates that the maximum pushout forces were consistent for Hole B, which may have resulted from fixture construction and the point of contact along the implant shaft. Additionally, data in this table suggest that the elastic limit of the setscrews was exceeded even at 10.9 N x cm, since pushout forces decreased when torque increased from 10.0 to 10.9 N x cm for Holes A and B, but this trend was not evident with Hole C and therefore these quadrants must be treated as outliers in the data set.

A multivariable linear regression model was fit to the outcome variable (mechanical pushout forces) as shown in Table 2. The significant predictor variables were screw torque (B = -13.40, p = 0.049), resonance frequency (B = -2.89, p = 0.012) and the interaction of screw torque and resonance frequency (B = 0.38, p = 0.004). Distance from the point of fixation, included in the model as a continuous variable, was not significant (B = -4.00, p = 0.36). The rationale for including distance as a continuous variable was that previous reports indicated that implant stability linearly decreases with increasing distances from the Smartpeg [55].

For verification of the linear reduction in resonance frequency ISQ values with increasing distance from the Smartpeg, a second model was fitted and distance was included in the model as dummy variables (Table 3). The first dummy variable was located 20 mm from the Smartpeg (Hole A) and the second dummy variable was located 30 mm from the Smartpeg (Hole C). The third position on the testing fixture was located 25 mm from the Smartpeg (Hole B) and acted as the reference point for the two dummy variables, since it was centrally located and the midpoint between fixation points (Figure 1). The significant predictor variables in this model were resonance frequency (B = -2.92, p = 0.008), interaction between resonance frequency and screw torque (B = 0.35, p = 0.005), and distance from Hole C to Hole B (B = 0.35, p = 0.005). However, the distance from Hole A to Hole B was not significant (p = 0.12). The predictor variable screw torque had a significant trend in which increased torques resulted in higher ISQ values and added to the interaction term, but this was not statistically significant alone (p = 0.07). A positive correlation between resonance frequency ISQ values and screw torque has also been previously reported in an in vitro model [60] and demonstrates the reproducibility of our work with previous studies.

The relationship of resonance frequency and screw torque to mechanical pushout forces is graphically represented in Figure 3 and quantitatively described in Table 4 using the same raw data. The three-dimensional representation shows a general increase in the force required to displace the titanium implant as screw torque and resonance frequency increased. The majority of the data points collected in the model occurred at higher ISQ values (37/45), which is a trend also noted in previous dental applications [48]. For ease of understanding of the relationship, resonance frequency was subdivided into ranges of low, low-to-medium, medium, and high values, which were used to predict the degree of osseointegration. The mean and SD of each quadrant are presented in Table 4.

DISCUSSION

The prolonged recovery and rehabilitation period that currently follows osseointegration implantation (approximately 12 months from postoperative to full weight bearing [73]) in individuals with lower-limb amputations demonstrates the need for a simple, noninvasive tool that determines implant strength and stability. The use of an RFA-type of rehabilitation device is anticipated to quantitatively determine the integrity of the bone-implant interface during osseointegration and increase the loading rate on osseointegrated implants, thereby shortening current rehabilitation regimens. Therefore, the aim of this study was to determine whether a relationship existed between resonance frequency and mechanical pushout forces in a carefully controlled in vitro medullary canal osseointegration model.

[FIGURE 3 OMITTED]

The correlation between screw torque and resonance frequency, as demonstrated in Figure 3, confirms the reliability of the Osstell device for predicting mechanical pushout forces in a nonuniform osseointegration model. While several data points were in the low and low-to-medium ranges, the data collected were primarily in the medium and high ranges of ISQ values and fit the published values for stable osseointegration ISQs ranging from 40 to 70 [48]. The majority of data collected in the model was in the ISQ range of 50 to 70, and the torques tested were in the vicinity of previous in vivo dental applications. While the exact ISQ value for complete osseointegration is unknown, Zix et al. report an average ISQ of 57.66 [+ or -] 8.19 (ISQ range: 23-73) [48], Kessler-Liechti et al. report 64.50 [+ or -] 7.90 (ISQ range: 58-72) [53], and Zhou et al. report 53.90 [+ or -] 7.70 (ISQ range: 37-68) [46] in in vivo human and rabbit models. These values are in the locality of the overall mean and SD in this experiment, which was 55.74 [+ or -] 7.35.

Since previous literature by Friberg et al. has shown that the density of the host bone bed and bone quality are factors affecting resonance frequency reporting [72], we selected an ABS material with an approximate density of 1.05 g/[cm.sup.3]. While this material resulted in a moderate to large coefficient of determination when screw torque, resonance frequency, fixation location, and the interaction between screw torque and resonance frequency were compared with mechanical pushout forces (Tables 2-3), the material type selected did not entirely replicate the density of cortical bone, which is reported to be 2.06 g/[cm.sup.3] in human femurs [74]. The discrepancy in material selection may have accounted for the adjusted coefficients of determination of 0.67 in the multiple linear regressions, which are considered to be a moderate-to-high-positive correlation [75], but may have been even larger if a different material type had been selected to help validate the model. However, a positive correlation between ISQ values and screw torque has been previously reported in an in vitro model and demonstrates the consistency of our model with previous studies [60]. Further investigation using fresh cadaveric bone samples is planned in the future and will serve to validate our current model.

In our experiment, the force required for implant removal was affected by the points of fixation, as indicated by significance between Hole B and Hole C, respectively. While a statistical relationship was not found for all hole locations, varying degrees of implant stability based on the distance of the Smartpeg to the proximal bone bed surface have been reported previously [55]. According to manufacturer specifications, "a change of about 3 ISQ/mm should be expected if implants are placed in the same bone density" [55]. However, investigation of Table 1 indicates that this trend was not observed in our model and may be due to implant geometry, which has been reported as a factor affecting stiffness in dental and orthopedic implants [60,76]. The hexagonal head of the implant allowed the Smartpeg to stay within the same distance to the ABS material, but the fixation point differed in 5 mm increments between hole locations. The rationale for including the hexagonal head in the model was that it allowed for reproducibility by preventing implant migration which would lead to confounding variables. However, there is reason to believe that a threaded design may have also allowed accurate data collection but would have required fabrication of multiple ABS constructs and screws because of anticipated damage to the screw threads during pushout tests.

While the data collected in the model was initially targeted at proving safety and efficacy in long bone medullary canals, high correlations between resonance frequency and mechanical pushout forces demonstrate the utility of the device for future use in short bones (digits) and low load-bearing implants (facial reconstruction). In all these cases, the biomechanical behavior of osseointegrated implants is vital for firm skeletal attachment [77], since stress must be applied gradually over time to prevent over- or underloading. In addition, to further reduce the likelihood of osseointegration failure due to high interfacial shear stresses, this model may provide useful boundary conditions for finite element models using our mechanical pushout forces as primary stability estimations.

Limitations of the ABS construct used in our model include the restrictions of torques exceeding 10.9 N x cm because of the elastic limit of the setscrews and the inability to determine the effect of resonance frequency on mechanical pushout with primary and secondary implant stability. The relationship determined in this study demonstrated a strong positive correlation between resonance frequency ISQ values and mechanical pushout forces but requires in vitro cadaveric validation and in vivo experimentation in long bones to monitor the degree of osseointegration over time with biomechanical testing.

CONCLUSIONS

The relationship between resonance frequency ISQ values and the force required for implant removal in a nonuniform bone-implant interface model had not been evaluated before this investigation. Orthopedic implants in human medullary canals often result in a nonuniform bone-implant interface [78], and this limits the speed of rehabilitation for individuals with limb loss. While our investigation used an in vitro testing modality, a clear distinction was evident from multiple linear regressions, which demonstrated that the interaction between resonance frequency and screw torque correlated with mechanical pushout forces. The simulated approach allowed for controlled loading at separate contact points to represent partial osseointegration, reaffirmed the utility of resonance frequency as an advocated nondestructive mechanical assessment of skeletal fixation, and demonstrated a direct relationship with implant displacements. While the exact force required for implant removal may not be necessary in dental applications, based on our findings in this in vitro model, RFA appears to hold promise for application as a rehabilitation tool for determining implant strength and stability following osseointegration implant placement, and this technology deserves further human cadaveric and in vivo investigation.

ACKNOWLEDGMENTS

Author Contributions:

Study concept and design: T. G. Rosenbaum Chou, K. N. Bachus.

Acquisition of data: B. M. Isaacson, R. E. Vance.

Analysis and interpretation of data: B. M. Isaacson, R. E. Vance.

Drafting of manuscript: B. M. Isaacson.

Critical revision of manuscript for important intellectual content: K. N. Bachus, R. D. Bloebaum, J. B. Webster.

Statistical analysis: B. M. Isaacson.

Obtained funding: J. B. Webster.

Administrative, technical, or material support: R. D. Bloebaum. Study supervision: J. B. Webster.

Financial Disclosures: The Osstell device and Smartpegs used in the study were donated by Anders Petersson, Chief Executive Officer of Osstell AB.

Funding/Support: This material was based on work supported (or supported in part) by the Department of Veterans Affairs (VA) Office of Research and Development, Rehabilitation Research and Development Service, VA Salt Lake City Health Care System, Salt Lake City, UT; the Albert & Margaret Hofmann Chair and the Department of Orthopaedics, University of Utah School of Medicine, Salt Lake City, UT; and University of Utah Center for Rehabilitation Research (grant CRR-06-PMR), Department of Physical Medicine and Rehabilitation, University of Utah, Salt Lake City, UT.

Additional Contributions: We would like to thank Gregory Stoddard for assistance with and interpretation of statistical analysis; Gwenevere Shaw for support with manuscript preparation; and Anders Petersson, Chief Executive Officer of Osstell AB, for generously providing an Osstell Mentor and Smartpegs for the experiment.

Submitted for publication June 19, 2009. Accepted in revised form October 14, 2009.

REFERENCES

[1.] Albrektsson T, Albrektsson B. Osseointegration of bone implants. A review of an alternative mode of fixation. Acta Orthop Scand. 1987;58(5):567-77. [PMID: 3321881]

[2.] Branemark PI. Osseointegration and its experimental background. J Prosthet Dent. 1983;50(3):399-410. [PMID: 6352924] DOI:10.1016/S0022-3913(83)80101-2

[3.] Branemark R, Branemark PI, Rydevik B, Myers RR. Osseointegration in skeletal reconstruction and rehabilitation: A review. J Rehabil Res Dev. 2001;38(2):175-81. [PMID: 11392650]

[4.] Pitkin M. On the way to total integration of prosthetic pylon with residuum. J Rehabil Res Dev. 2009;46(3):345-60. [PMID: 19675987] DOI:10.1682/JRRD.2008.08.0112

[5.] Staubach KH, Grundei H. [The first osseointegrated percutaneous prosthesis anchor for above-knee amputees]. Biomed Tech (Berl). 2001;46(12):355-61. German. [PMID: 11820163] DOI:10.1515/bmte.2001.46.12.355

[6.] Branemark PI, Grondahl HG, Branemark BK. Why osseointegration would work and how it did in the first patients treated. Basic facts and philosophical thoughts. In: Branemark PI, editor. The osseointegration book: From calvarium to calcaneus. Berlin (Germany): Quintessence; 2005.

[7.] Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plas Reconstr Surg. 1969; 3(2):81-100. [PMID: 4924041] DOI:10.3109/02844316909036699

[8.] Jemt T. Failures and complications in 391 consecutively inserted fixed prostheses supported by Branemark implants in edentulous jaws: A study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants. 1991;6(3):270-76. [PMID: 1813395]

[9.] Kegel B, Carpenter ML, Burgess EM. Functional capabilities of lower extremity amputees. Arch Phys Med Rehabil 1978;59(3):109-20. [PMID: 646596]

[10.] Rommers GM, Vos LD, Groothoff JW, Eisma WH. Mobility of people with lower limb amputations: Scales and questionnaires: A review. Clin Rehabil. 2001;15(1):92-102. [PMID: 11237166] DOI:10.1191/026921501677990187

[11.] Aaron RK, Herr HM, Ciombor DM, Hochberg LR, Donoghue JP, Briant CL, Morgan JR, Ehrlich MG. Horizons in prosthesis development for the restoration of limb function. J Am Acad Orthop Surg. 2006;14(10 Spec No.):S198-204. [PMID: 17003199]

[12.] Hagberg K, Branemark R. Consequences of non-vascular trans-femoral amputation: A survey of quality of life, prosthetic use and problems. Prosthet Orthot Int. 2001;25(3): 186-94. [PMID: 11860092] DOI:10.1080/03093640108726601

[13.] Isaacson BM, Stinstra JG, MacLeod RS, Webster JB, Beck JP, Bloebaum RD. Bioelectric analyses of an osseointe grated intelligent implant design system for amputees. J Vis Exp. 2009(29):1-6. [PMID: 19609251]

[14.] Albrektsson T, Branemark PI, Hansson HA, Lindstrom J. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta Orthop Scand. 1981;52(2):155-70. [PMID: 7246093]

[15.] Meyer U, Joos U, Mythili J, Stamm T, Hohoff A, Fillies T, Stratmann U, Wiesmann HP. Ultrastructural characterization of the implant/bone interface of immediately loaded dental implants. Biomaterials. 2004;25(10):1959-67. [PMID: 14738860] DOI:10.1016/j.biomaterials.2003.08.070

[16.] Hofmann AA, Bloebaum RD, Bachus KN. Progression of human bone ingrowth into porous-coated implants. Rate of bone ingrowth in humans. Acta Orthop Scand. 1997;68(2): 161-66. [PMID: 9174454]

[17.] Slaets E, Naert I, Carmeliet G, Duyck J. Early cortical bone healing around loaded titanium implants: A histological study in the rabbit. Clin Oral Implants Res. 2009;20(2): 126-34. [PMID: 19191791] DOI:10.1111/j.1600-0501.2008.01623.x

[18.] Duyck J, Vandamme K, Geris L, Van Oosterwyck H, De Cooman M, Vandersloten J, Puers R, Naert I. The influence of micro-motion on the tissue differentiation around immediately loaded cylindrical turned titanium implants. Arch Oral Biol. 2006;51(1):1-9. [PMID: 15922992] DOI:10.1016/j.archoralbio.2005.04.003

[19.] Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH. Timing of loading and effect of micromotion on bonedental implant interface: Review of experimental literature. J Biomed Mater Res. 1998;43(2):192-203. [PMID: 9619438] DOI:10.1002/(SICI)1097-4636(199822)43:2<192::AIDJBM14>3.0.CO;2-K

[20.] Glauser R, Sennerby L, Meredith N, Ree A, Lundgren A, Gottlow J, Hammerle CH. Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants. Clin Oral Implants Res. 2004;15(4):428-34. [PMID: 15248877] DOI:10.1111/j.1600-0501.2004.01036.x

[21.] Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res. 1996;7(3):261-67. [PMID: 9151590] DOI:10.1034/j.1600-0501.1996.070308.x

[22.] Roser K, Johansson CB, Donath K, Albrektsson T. A new approach to demonstrate cellular activity in bone formation adjacent to implants. J Biomed Mater Res. 2000;51(2): 280-91. [PMID: 10825228] DOI:10.1002/(SICI)1097-4636(200008)51:2<280::AIDJBM17>3.0.CO;2-L

[23.] Nkenke E, Lehner B, Weinzierl K, Thams U, Neugebauer J, Steveling H, Radespiel-Troger M, Neukam FW. Bone contact, growth, and density around immediately loaded implants in the mandible of mini pigs. Clin Oral Implants Res. 2003;14(3):312-21. [PMID: 12755781] DOI: 10.1034/j.1600-0501.2003.120906.x

[24.] Slaets E, Carmeliet G, Naert I, Duyck J. Early trabecular bone healing around titanium implants: A histologic study in rabbits. J Periodontol. 2007;78(3):510-17. [PMID: 17335375] DOI: 10.1902/jop.2007.060183

[25.] Piattelli A, Corigliano M, Scarano A, Costigliola G, Paolantonio M. Immediate loading of titanium plasma-sprayed implants: An histologic analysis in monkeys. J Periodontol. 1998;69(3):321-27. [PMID: 9579618]

[26.] Vandamme K, Naert I, Vander Sloten J, Puers R, Duyck J. Effect of implant surface roughness and loading on periimplant bone formation. J Periodontal. 2008;79(1):150-57. [PMID: 18166105] DOI: 10.1902/jop.2008.060413

[27.] Pefiarrocha M, Boronat A, Garcia B. Immediate loading of immediate mandibular implants with a full-arch fixed prosthesis: A preliminary study. J Oral Maxillofac Surg. 2009; 67(6):1286-93. [PMID: 19446218] DOI: 10.1016/j.joms.2008.12.024

[28.] Abrahamsson I, Linder E, Lang NP. Implant stability in relation to osseointegration: An experimental study in the Labrador dog. Clin Oral Implants Res. 2009;20(3):313-18. [PMID: 19405177] DOI:10.1111/j.1600-0501.2008.01646.x

[29.] Seballe K, Hansen ES, B-Rasmussen H, Jergensen PH, Bunger C. Tissue ingrowth into titanium and hydroxyapatite-coated implants during stable and unstable mechanical conditions. J Orthop Res. 1992;10(2):285-99. [PMID: 1311039] DOI:10.1002/jor. 1100100216

[30.] Adell R, Hansson BO, Branemark PI, Breine U. Intraosseous anchorage of dental prostheses. II. Review of clinical approaches. Scand J Plast Reconstr Surg. 1970;4(1):19-34. [PMID: 4920517] DOI: 10.3109/02844317009038440

[31.] Rafel SS. Temperature changes during high-speed drilling on bone. J Oral Surg Anesth Hosp Dent Serv. 1962;20: 475-77. [PMID: 13990417]

[32.] Taylor TD. Prosthodontic problems and limitations associated with osseointegration. J Prosthet Dent. 1998;79(1):74-78. [PMID: 9474545] DOI:10.1016/S0022-3913(98)70197-0

[33.] Bloebaum RD, Rubman MH, Hofmann AA. Bone ingrowth into porous-coated tibial components implanted with autograft bone chips: Analysis of ten consecutively retrieved implants. J Arthroplasty. 1992;7(4):483-93. [PMID: 1479367]

[34.] Pilliar RM, Lee JM, Maniatopoulos C. Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res. 1986;(208):108-13. [PMID: 3720113]

[35.] Kieswetter K, Schwartz Z, Dean DD, Boyan BD. The role of implant surface characteristics in the healing of bone. Crit Rev Oral Biol Med. 1996;7(4):329-45. [PMID: 8986395] DOI:10.1177/10454411960070040301

[36.] Branemark PI, Albrektsson T. Titanium implants permanently penetrating human skin. Scand J Plast Reconstr Surg. 1982;16(1):17-21. [PMID: 7112035] DOI:10.3109/02844318209006565

[37.] Bloebaum RD, Bachus KN, Momberger NG, Hofmann AA. Mineral apposition rates of human cancellous bone at the interface of porous coated implants. J Biomed Mater Res. 1994;28(5):537-44. [PMID: 8027094] DOI:10.1002/jbm.820280503

[38.] Bloebaum RD, Bachus KN, Rubman MH, Dorr LD. Postmortem comparative analysis of titanium and hydroxyapatite porous coated femoral implants retrieved from the same patient. J Arthroplasty. 1993;8(2):203-11. [PMID: 8386749] DOI:10.1016/S0883-5403(09)80014-4

[39.] Vandamme K, Naert I, Geris L, Sloten JV, Puers R, Duyck J. Histodynamics of bone tissue formation around immediately loaded cylindrical implants in the rabbit. Clin Oral Implants Res. 2007;18(4):471-80. [PMID: 17517061] DOI:10.1111/j.1600-0501.2007.01339.x

[40.] Uhthoff HK, Germain JP. The reversal of tissue differentiation around screws. Clin Orthop Relat Res. 1977;(123): 248-52. [PMID: 856516]

[41.] Hagberg K, Branemark R, Gunterberg B, Rydevik B. Osseointegrated trans-femoral amputation prostheses: Prospective results of general and condition-specific quality of life in 18 patients at 2-year follow-up. Prosthet Orthot Int. 2008;32(1):29-41. [PMID: 18330803] DOI:10.1080/03093640701553922

[42.] Frossard L, Lee Gow D, Contoyannis B, Ewins D, Sullivan J, Tranberg R, Haggstrom E, Branemark R. Loading applied to the implant of transfemoral amputees fitted with a direct skeletal fixation during load bearing exercises. Fourth Australasian Biomechanics Conference; 2002 Nov 28-30; Melbourne, Australia. Melbourne (Australia): La Trobe University; 2002.

[43.] Bloebaum RD, Willie BM, Mitchell BS, Hofmann AA. Relationship between bone ingrowth, mineral apposition rate, and osteoblast activity. J Biomed Mater Res A. 2007; 81(2):505-14. [PMID: 17236212] DOI:10.1002/jbm.a.31087

[44.] Oh JS, Kim SG, Lim SC, Ong JL. A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;107(4):513-18. [PMID: 19095473] DOI:10.1016/j.tripleo.2008.08.026

[45.] Slaets E, Carmeliet G, Naert I, Duyck J. Early cellular responses in cortical bone healing around unloaded titanium implants: An animal study. J Periodontol. 2006;77(6): 1015-24. [PMID: 16734577] DOI:10.1902/jop.2006.050196

[46.] Zhou Y, Jiang T, Qian M, Zhang X, Wang J, Shi B, Xia H, Cheng X, Wang Y. Roles of bone scintigraphy and resonance frequency analysis in evaluating osseointegration of endosseous implant. Biomaterials. 2008;29(4):461-74. [PMID: 17983650] DOI:10.1016/j.biomaterials.2007.10.021

[47.] Andres-Garcia R, Vives NG, Climent FH, Palacin AF, Santos VR, Climent MH, Bullon P. In vitro evaluation of the influence of the cortical bone on the primary stability of two implant systems. Med Oral Patol Oral Cir Bucal. 2009; 14(2):E93-97. [PMID: 19179957]

[48.] Zix J, Lieger O, Iizuka T. Use of straight and curved 3-dimensional titanium miniplates for fracture fixation at the mandibular angle. J Oral Maxillofac Surg. 2007;65(9): 1758-63. [PMID: 17719394] DOI:10.1016/j.joms.2007.03.013

[49.] Atsumi M, Park SH, Wang HL. Methods used to assess implant stability: Current status. Int J Oral Maxillofac Implants. 2007;22(5):743-54. [PMID: 17974108]

[50.] Sunden S, Grondahl K, Grondahl HG. Accuracy and precision in the radiographic diagnosis of clinical instability in Branemark dental implants. Clin Oral Implants Res. 1995;6(4):220-26. [PMID: 8603113] DOI:10.1034/j.1600-0501.1995.060404.x

[51.] Sartoris DJ, Resnick D. Current and innovative methods for noninvasive bone densitometry. Radiol Clin North Am. 1990;28(2):257-78. [PMID: 2408094]

[52.] Sennerby L, Meredith N. Implant stability measurements using resonance frequency analysis: Biological and biomechanical aspects and clinical implications. Periodontol 2000. 2008;47:51-66. [PMID: 18412573] DOI: 10.1111/j.1600-0757.2008.00267.x

[53.] Kessler-Liechti G, Zix J, Mericske-Stern R. Stability measurements of 1-stage implants in the edentulous mandible by means of resonance frequency analysis. Int J Oral Maxillofac Implants. 2008;23(2):353-58. [PMID: 18548934]

[54.] Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Implants Res. 1997;8(3):234-43. [PMID: 9586468] DOI:10.1034/j.1600-0501.1997.080310.x

[55.] Turkyilmaz I, Sennerby L, Yilmaz B, Bilecenoglu B, Ozbek EN. Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: A human cadaver study. Clin Implant Dent Relat Res. 2009;11(1):52-58. [PMID: 18384400] DOI: 10.1111/j.1708-8208.2008.00095.x

[56.] Balshi SF, Wolfinger GJ, Balshi TJ. An examination of immediately loaded dental implant stability in the diabetic patient using resonance frequency analysis (RFA). Quintessence Int. 2007;38(4):271-79. [PMID: 17432781]

[57.] Boronat Lopez A, Balaguer Martinez J, Lamas Pelayo J, Carrillo Garcia C, Pefiarrocha Diago M. Resonance frequency analysis of dental implant stability during the healing period. Med Oral Patol Oral Cir Bucal. 2008;13(4): E244-47. [PMID: 18379449]

[58.] Rao W, Benzi R. Single mandibular first molar implants with flapless guided surgery and immediate function: Preliminary clinical and radiographic results of a prospective study. J Prosthet Dent. 2007;97(6 Suppl):S3-S14. [PMID: 17618931] DOI:10.1016/S0022-3913(07)60003-1

[59.] Ratner BD, Hoffman AS, Schoen FJ, Lemons JE. Biomaterials science: An introduction to materials in medicine. 2nd ed. Boston (MA): Elsevier Academic Press; 2004. p. 543.

[60.] Tozum TF, Turkyilmaz I, McGlumphy EA. Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects: An in vitro study. J Oral Rehabil. 2008;35(10): 739-44. [PMID: 18422511] DOI:10.1111/j.1365-2842.2007.01840.x

[61.] Capek L, Simunek A, Slezak R, Dzan L. Influence of the orientation of the Osstell transducer during measurement of dental implant stability using resonance frequency analysis: A numerical approach. Med Eng Phys. 2009;31(7): 764-69. [PMID: 19297232] DOI:10.1016/j.medengphy.2009.02.003

[62.] Kuo TY, Skedros JG, Bloebaum RD. Comparison of human, primate, and canine femora: Implications for biomaterials testing in total hip replacement. J Biomed Mater Res. 1998;40(3):475-89. [PMID: 9570081] DOI:10.1002/(SICI)1097-4636(19980605)40:3<475::AIDJBM19>3.0.CO;2-I

[63.] Bloebaum RD, Ota DT, Skedros JG, Mantas JP. Comparison of human and canine external femoral morphologies in the context of total hip replacement. J Biomed Mater Res. 1993;27(9):1149-59. [PMID: 8126013] DOI: 10.1002/jbm.820270905

[64.] Rosenbaum TG, Hamblin T, Bloebaum RD. Determining the degree of cortical bone asymmetry in bilateral, nonpathological, human femur pairs. J Biomed Mater Res A. 2006;76(3):450-55. [PMID: 16270336] DOI:10.1002/jbm.a.30568

[65.] Nyman JS, Roy A, Shen X, Acuna RL, Tyler JH, Wang X. The influence of water removal on the strength and toughness of cortical bone. J Biomech. 2006;39(5):931-38. [PMID: 16488231] DOI:10.1016/j.jbiomech.2005.01.012

[66.] Pospula W, Noor TA, Al Rowaih A. Cementless Zweymuller hip replacement: A short-term follow-up in Al Razi Hospital, Kuwait. Med Princ Pract. 2005;14(4):255-59. [PMID: 15961936] DOI:10.1159/000085745

[67.] Vervest TM, Anderson PG, Van Hout F, Wapstra FH, Louwerse RT, Koetsier JW. Ten to twelve-year results with the Zweymuller cementless total hip prosthesis. J Arthroplasty. 2005;20(3):362-68. [PMID: 15809956] DOI:10.1016/j.arth.2004.11.017

[68.] Beder OE, Eade G. An investigation of tissue tolerance to titanium metal implants in dogs. Surgery. 1956;39(3):470-73. [PMID: 13299000]

[69.] Emneus H, Gudmundsson G. Final report on the clinical testing of titanium. Acta Orthop Scand. 1967;38(1):372-73. Available from: http://pdfserve.informaworld.com/67884_768420410_784865168.pdf/.

[70.] Martin AD, Driscoll CL, Wood CP, Felmlee JP. Safety evaluation of titanium middle ear prostheses at 3.0 tesla. Otolaryngol Head Neck Surg. 2005;132(4):537-42. [PMID: 15806041] DOI: 10.1016/j.otohns.2004.12.003

[71.] Wichmann W, Von Ammon K, Fink U, Weik T, Yasargil GM. Aneurysm clips made of titanium: Magnetic characteristics and artifacts in MR. AJNR Am J Neuroradiol. 1997; 18(5):939-44. [PMID: 9159374]

[72.] Friberg B, Jemt T, Lekholm U. Early failures in 4,641 consecutively placed Branemark dental implants: A study from stage 1 surgery to the connection of completed prostheses. Int J Oral Maxillofac Implants. 1991;6(2):142-46. [PMID: 1809668]

[73.] Ward DA, Robinson KP. Osseointegration for the skeletal fixation of limb prostheses in amputations at the transfemoral level. In: Branemark PI, editor. The osseointegration book: From calvarium to calcaneus. Berlin (Germany): Quintessence; 2005. p. 463-76.

[74.] Currey JD. Mechanical properties of bone tissues with greatly differing functions. J Biomech. 1979;12(4):313-19. [PMID: 468855] DOI:10.1016/0021-9290(79)90073-3

[75.] Hinkle DE, Wiersma W, Jurs SG. Applied statistics for the behavioral sciences. 5th ed. Boston (MA): Houghton Miff lin; 2003.

[76.] Dhert WJ, Verheyen CC, Braak LH, De Wijn JR, Klein CP, De Groot K, Rozing PM. A finite element analysis of the push-out test: Influence of test conditions. J Biomech Meter Res. 1992;26(1):119-30. [PMID: 1577832] DOI:10.1002/jbm.820260111

[77.] Branemark R, Skalak R. An in-vivo method for biomechanical characterization of bone-anchored implants. Med Eng Phys. 1998;20(3):216-19. [PMID: 9690492] DOI:10.1016/S1350-4533(98)00023-X

[78.] Rosenbaum TG, Bloebaum RD, Ashrafi S, Lester DK. Ambulatory activities maintain cortical bone after total hip arthroplasty. Clin Orthop Relat Res. 2006;450:129-37. [PMID: 16721349] DOI:10.1097/01.blo.0000223981.21584.35

DOI: 10.1682/JRRD.2009.02.0012

Abbreviations: ABS = acrylonitrile butadiene styrene, ISQ = implant stability quotient, RFA = resonance frequency analysis, SD = standard deviation, VA = Department of Veterans Affairs.

Brad M. Isaacson, BS; (1-2) * Richard E. Vance, BS; (3) Teri G. Rosenbaum Chou, PhD; (4) Roy D. Bloebaum, PhD; (1-3) Kent N. Bachus, PhD; (1-3) Joseph B. Webster, MD (1,5)

(1) Bone and Joint Research Laboratory, Department of Veterans Affairs Salt Lake City Health Care System, Salt Lake City, UT; (2) Department of Bioengineering, University of Utah, Salt Lake City, UT; (3) Department of Orthopaedics, University of Utah, Salt Lake City, UT; (4) Orthocare Innovations, Seattle, WA; (5) Department of Physical Medicine and Rehabilitation, University of Utah, Salt Lake City, UT

* Address all correspondence to Brad M. Isaacson, BS; Bone and Joint Research Laboratory (151F), VA Salt Lake City Health Care System, Salt Lake City, UT 84148; 801-5821565, ext 4122; fax: 801-584-2533. Email: brad.isaacson@utah.edu
Table 1.

Mean [+ or -] standard deviation mechanical pushout forces
independently assessed as a function of screw torque and distance from
Smartpeg (n = 3 per combination, total N = 45).

               Distance irom smartpeg

Screw Torque
(N x cm)       2.0 mm, Hole A

5.0            26.83 [+ or -] 4.21
7.0            26.19 [+ or -] 8.74
8.6            41.60 [+ or -] 12.46
10.0           73.40 [+ or -] 0.00
10.9           58.82 [+ or -] 15.46

               Distance irom smartpeg

Screw Torque
(N x cm)       2.5 mm, Hole B

5.0            35.80 [+ or -] 3.76
7.0            46.13 [+ or -] 3.14
8.6            47.95 [+ or -] 9.55
10.0           79.90 [+ or -] 3.76
10.9           70.61 [+ or -] 18.10

               Distance irom smartpeg

Screw Torque
(N x cm)       3.0 mm, Hole C

5.0            33.37 [+ or -] 3.72
7.0            26.19 [+ or -] 4.15
8.6            47.95 [+ or -] 1.57
10.0           61.13 [+ or -] 4.21
10.9           68.79 [+ or -] 4.71

Table 2.

Multiple linear regression model of mechanical pushout (N = 45), using
distance as a continuous variable.

Predictor Variable      Coefficient (B)   Standard Error   p-Value

Distance                         -4.00             4.34      0.36
Screw Torque                    -13.40             6.60     0.049
Resonance Frequency              -2.89             1.10     0.012
Interaction Between               0.38             0.12     0.004
  Resonance Frequency
  and Screw Torque

Note: Multiple R = 0.83, multiple [R.sup.2] = 0.69, multiple adjusted
[R.sup.2] = 0.67.

Table 3.

Multiple linear regression model of mechanical pushout (N = 45), using
distance from the Smartpeg as dummy variables.

                                 Coefficient   Standard
Predictor Variable               (B)           Error      p-Value

Screw Torque                     -11.72         6.32       0.07
Resonance Frequency               -2.92         1.05       0.008
Interaction Between Resonance      0.35         0.12       0.005
  Frequency and Screw Torque
Distance from Hole A to Hole B    -7.10         4.51       0.12
Distance from Hole C to Hole B     0.35         0.12       0.005

Note: Multiple R = 0.84, multiple [R.sup.2] = 0.71, multiple adjusted
[R.sup.2] = 0.67.

Table 4.

Force (mean [+ or -] standard deviation) required for mechanical
pushout subdivided by resonance frequency and screw torques, where "--"
signifies that no data were collected in the quadrant (N = 45).

Screw Torque   Low, 30-40        Low-Medium, 40-50
(N x cm)       (2 data points)   (6 data points)

5.0            --                34.8 [+ or -] 4.2
7.0            --                --
8.6            30.7              --
10.0           --                58.7
10.9           41.6              71.5

                              ISQ Range

Screw Torque   Medium, 50-60       High, 60-70
(N x cm)       (25 data points)    (12 data points)

5.0            29.8 [+ or -] 5.3   --
7.0            28.0 [+ or -] 7.0   36.7 [+ or -] 13.1
8.6            47.6 [+ or -] 5.9   47.9 [+ or -] 9.6
10.0           73.1 [+ or -] 7.8   --
10.9           67.9 [+ or -] 5.7   69.5 [+ or -] 15.0

ISQ = implant stability quotient.